Constant Volume Sample Cutter
Volumetric Die Cutter for the preparation of samples for MD Rheometer, OD Rheometer and Mooney Viscometer


Standards the instrument complies with:   
ASTM D1646;   ASTM D2084;   ASTM D5289;   ISO 289-1;   ISO 289-2;   ISO 289-4;   ISO 6502-1;   ISO 6502-2;   ISO 6502-3;    

Overview 
Pneumatic cutter for the preparation of constant-volume samples for MD Rheometer, OD Rheometer and Mooney Viscometer from polymer or green rubber compounds.
The solid construction of the machine and the double-piston system permit to obtain constant volume samples even with hard and high viscosity compounds.
By using samples with constant volume for your rheology tests you will increase test repeatability ensuring correct filling of test chamber.


Key Features 
• Exclusive adjustment system, which permits to set the machine to obtain samples for:
- Moving Die Rheometer
- Oscillating Disk Rheometer
- Mooney Viscometer
• Puncher for obtaining the central hole in the lower sample for the Mooney viscosity test
• Pneumatically-activated safety panel
• CE Labelling


Select the Type of Sample 
The selector, located on the back of the instrument, allows you to select the type of specimen to be produced.



Sample Volume for MD Rheometer 
According to ASTM D 5289: "The test specimen is considered to be of proper size when a small bead of rubber compound is extruded uniformly around the periphery of the dies as they are closed(116 to 160% of the test cavity volume).
For an instrument with a test cavity volume of 3.44 cm3, this is achieved when the test specimen volume is between 4.0 and 5.5 cm3.
Undersized test specimens can cause low cavity pressure and low torque readings. Oversized test specimens cool the dies excessively during the early part of the test period, affecting the vulcanization characteristics."

The standard setup of Gibitre Volumetric cutter produces each sample for MD Rheometer test with 4.8 cm3 volume.
At order time you can request a custom setup to produce samples with a volume between 4 and 5.5 cm3.

Sample Volume for Mooney Viscometer 
According to ASTM D1646 : "The test specimen shall consist of two test pieces of the material being tested having a combined volume of 25 cm3. This volume is approximately 1.5 times the volume of the test cavity (1.45 times for small rotor, 1.67 times for large rotor) and will ensure that the cavity is completely filled."

The standard setup of Gibitre Volumetric cutter produces each sample for mooney test with 12.5 cm3.
At order time you can request a custom setup to produce samples with a volume between 12 and 13 cm3.

Puncher for Hole for Mooney Sample  
A puncher, incorporated in the instrument, permits the central hole to be made in the lower sample for the Mooney viscosity test.
The pneumatically-activated safety panel ensures the conformity of the instrument with CE safety requirements.




Safety devices 
The instrument is equipped with:
• Pneumatic activation of the safety protection screen
• Safety switch, which prevents the activation of the piston if the Safety Protection Screen is not closed.
• CE Labelling


Development and production 
The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.



Instrument Characteristics 
 
Samples obtainable 
Moving Die Rheometer: 4.0 to 5.5 cm3 (default 4.8 cm3)
Mooney Viscometer (per sample): 12 to 13 cm3 (default 12.5 cm3).
Oscillating Disk Rheometer: 8 to 11 cm3 (default 9.5 cm3).
Note: non-default volumes to be requested at order 
Selection of the type of sample 
The Selector in the back part of the instrument permits to select the type of sample to produce 
Force produced 
1200 Kg (at six bar air pressure) 
Safety Devices 
 
Safety devices 
Transparent protection door with safety switch 
Labelling 
CE Labelling 
Technical specifications 
 
Compressed air line pressure 
6 bar (min.) 
Compressed air consumption 
Approx.14 NL/cycle 
Instrument overall dimensions 
(Depth x Width x Height) 400 x 400 x 1000 mm 
Instrument weight 
80 Kg 


Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Volumetric Die Cutter for the preparation of samples for MD Rheometer, OD Rheometer and Mooney Viscometer

Constant Volume Sample Cutter

Get a Quotation
Volumetric Die Cutter for the preparation of samples for MD Rheometer, OD Rheometer and Mooney Viscometer
Why Choose 'Constant Volume Sample Cutter'?
Overview 
Pneumatic cutter for the preparation of constant-volume samples for MD Rheometer, OD Rheometer and Mooney Viscometer from polymer or green rubber compounds.
The solid construction of the machine and the double-piston system permit to obtain constant volume samples even with hard and high viscosity compounds.
By using samples with constant volume for your rheology tests you will increase test repeatability ensuring correct filling of test chamber.
Key Features 
• Exclusive adjustment system, which permits to set the machine to obtain samples for:
- Moving Die Rheometer
- Oscillating Disk Rheometer
- Mooney Viscometer
• Puncher for obtaining the central hole in the lower sample for the Mooney viscosity test
• Pneumatically-activated safety panel
• CE Labelling
Select the Type of Sample 
The selector, located on the back of the instrument, allows you to select the type of specimen to be produced.
 
Sample Volume for MD Rheometer 
According to ASTM D 5289: "The test specimen is considered to be of proper size when a small bead of rubber compound is extruded uniformly around the periphery of the dies as they are closed(116 to 160% of the test cavity volume).
For an instrument with a test cavity volume of 3.44 cm3, this is achieved when the test specimen volume is between 4.0 and 5.5 cm3.
Undersized test specimens can cause low cavity pressure and low torque readings. Oversized test specimens cool the dies excessively during the early part of the test period, affecting the vulcanization characteristics."

The standard setup of Gibitre Volumetric cutter produces each sample for MD Rheometer test with 4.8 cm3 volume.
At order time you can request a custom setup to produce samples with a volume between 4 and 5.5 cm3.
Sample Volume for Mooney Viscometer 
According to ASTM D1646 : "The test specimen shall consist of two test pieces of the material being tested having a combined volume of 25 cm3. This volume is approximately 1.5 times the volume of the test cavity (1.45 times for small rotor, 1.67 times for large rotor) and will ensure that the cavity is completely filled."

The standard setup of Gibitre Volumetric cutter produces each sample for mooney test with 12.5 cm3.
At order time you can request a custom setup to produce samples with a volume between 12 and 13 cm3.
Puncher for Hole for Mooney Sample  
A puncher, incorporated in the instrument, permits the central hole to be made in the lower sample for the Mooney viscosity test.
The pneumatically-activated safety panel ensures the conformity of the instrument with CE safety requirements.
Safety devices 
The instrument is equipped with:
• Pneumatic activation of the safety protection screen
• Safety switch, which prevents the activation of the piston if the Safety Protection Screen is not closed.
• CE Labelling
Development and production 
The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.
Technical specification
Reference Standards
Standards the instrument complies with:  ISO 289-1; ISO 289-2; ISO 289-4; ASTM D1646; ASTM D2084; ASTM D5289; ISO 6502-1; ISO 6502-2; ISO 6502-3;  
Instrument Characteristics
Samples obtainable  Moving Die Rheometer: 4.0 to 5.5 cm3 (default 4.8 cm3)
Mooney Viscometer (per sample): 12 to 13 cm3 (default 12.5 cm3).
Oscillating Disk Rheometer: 8 to 11 cm3 (default 9.5 cm3).
Note: non-default volumes to be requested at order 
Selection of the type of sample  The Selector in the back part of the instrument permits to select the type of sample to produce  Force produced  1200 Kg (at six bar air pressure) 
Safety Devices
Safety devices  Transparent protection door with safety switch  Labelling  CE Labelling 
Technical specifications
Compressed air line pressure  6 bar (min.)  Compressed air consumption  Approx.14 NL/cycle  Instrument overall dimensions  (Depth x Width x Height) 400 x 400 x 1000 mm  Instrument weight  80 Kg 
References
Gibitre Constant Volume Sample Cutter
is currently installed in
23 countries
Request the List of top References
Download
Quick brochure

   
Product leaflet

General catalogue

CE certificate
example
News
Call to action
Ask for quotation
a guided way to ask
for a quotation
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
Your calls are always welcome
monday to friday
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.