Laboratory Press - Drive
Programmable Laboratory press for compression moulding, with 270x270 mm platens and 25 T closure force


Standards the instrument complies with:   
ASTM D1928;   ASTM D3182;   ASTM D4703-16;   EN 205;   ISO 293;   ISO 2393;   ISO 16770;   ISO 6916-1;   ISO 6916-2;   UNI 5572;    

 
Laboratory press for preparing plates and test specimens compliant with international standards, engineered for uniform heat distribution on the platens and high mechanical rigidity. The result is specimens with consistent properties and thickness, ready for measurement tools and materials quality control.

Ideal for producing specimens to established practices, including compression of thermoplastics (ISO 293, ASTM D1928, ASTM D4703-16) and elastomeric compounds (ISO 2393, ASTM D3182), as well as plates for tests such as ISO 16770 and ISO 6916-1.

A four-column frame with self-lubricating bushings ensures precise alignment and stability; flat heating elements maintain even temperature. A PLC with 10'' touchscreen manages up to 40 automatic cycles, boosting repeatability and productivity. Safety and cleanliness are ensured by an interlocked pneumatic door and enclosed area with fume extraction, CE marked.

Accessories broaden applications: controlled cooling for thermoplastics, phase-b

Overview 
The laboratory press is an important tool for the preparing of samples with reproducible characteristics.
Uniform temperature distribution over the platen Surface and mechanical solidity to ensure constant thickness of the sample and consistent closure force are basic elements to ensure this result.


Key Features 
• Developed and Produced in Italy
• Fully conforming to ISO 2393, ISO 293, ASTM D3182 Standards
• Mechanical structure with 4 columns (with 60 mm diameter) 
• Hydraulic system with 25 tons closure force
• 270x270 mm chrome plated steel platens 
• Extremely uniform temperature distribution over the platen surface (± 0.5°C)
• Hi speed heating (up to 15°C/min) and Cooling (up to 50°C/min)
• Touch-Screen display for complete machine control
• 40 programmable moulding cycles
• Connection to software for storage to SQL database of the moulding cycles & curves of your products
• Closed moulding area for easy fume aspiration
• Pneumatic Protection doors with safety lock
• CE Labelling




Tiltable Press Control Display 
The control of all the machine functions is carried out through the Touch-Screen Display integrated in the front part.
The 10" color display has been developed to ensure ergonomic use by the operator: it can be tilted during programming operations and positioned in a vertical position so as not to hinder access to the molding area.


Press Control Functions 
The PLC controller for the press is provided with a user-friendly Touch Screen Display and permits the free definition of the moulding cycles by setting:
• The displacements of the moving platen of the press,
• The thermal cycle (different temperatures of the platens and thermal gradients can be set within a moulding cycle)
• The pressure cycle (with the force control option installed).
40 Moulding cycles can be stored. Each cycle includes 30 moulding steps.


Gibitre_Press_Control Software: complete traceability of the sample preparation 
The press control PLC can be connected (via USB connection) to the Gibitre_Press_Control software.
The software allows you to:
• Prepare the list of products to be moulded
• Automatically send the Pre-Set temperature, for the product to be moulded, to the PLC of the press and activate its automatic adjustment
• Automatically send the test cycle required, for the product to be moulded, to the press PLC
• Display the trend curves of the platen temperatures and the closing force (if the option is installed) during the execution of the molding cycle.
• Store curves and numerical data in the Gibitre SQL database to keep the complete traceability of the specimen preparation cycle.
Gibitre_Press_Control software is compatible with Windows 10 and 11.




Closure Force Control 
The Digital pressure regulation system is designed for applications that require an accurate control of closure force.
A specific 2-stages hydraulic circuit, which includes proportional regulation valves and continuous fine regulation pumping system, has been designed to ensure accurate pressure control.
The results is a smooth and consistent force control which permit to set the closure force between 10 kN and 250 kN with 10 N set force regulation steps with regulation oscillations lower than 1000 N.
The control of the closing force is performed using the Automatic Press Control that must be installed in combination with the Closure Force Control option.
The Automatic Press Control device permits to set multiple pressure regulation steps within the moulding cycle.


Temperarture of the Platens 
The heating of the press surfaces is carried out by means of flat heating elements specially designed to meet the requirements of the ISO 2393 standard which admits a maximum difference of ± 0.5 °C inside the surface of the platens.
The temperature calibration is carried out through a rigorous control procedure which involves the measurement of 10 points distributed within the molding area.

The heating elements, with a power of 3.5 kW each, allow you to set heating gradients of the platens up to 15 °C/min.
Note: the maximum speed is conditioned by the thermal mass of the mold and the material to be molded.


Cooling system with Water circuit 
The cooling of the press platens during the molding cycle is necessary for the molding of thermoplastic materials because the material must be cooled down to the solidification temperature before opening the mold.

Gibitre presses can be equipped with an efficient cooling device that allows the temperature reduction of the platens at a controlled speed (down to -50°C/min).

The cooling device consists of a cooling circuit that can be connected to a demineralised water supply source or to a chiller.

The cooling speed is set in the test cycle and is modulated by the control PID independently in the two planes of the press.




Cooling system with Compressed Air circuit 
Gibitre presses can be equipped with a cooling device by means of compressed air which allows the temperature reduction of the platens at a controlled speed (down to -3 °C/min).

The device allows to speed up the regulation of the press temperature when molding operations at different temperatures have to be carried out.

The cooling speed is set in the test cycle and is modulated by the control PID independently in the two planes of the press.


Light panel 
The press is equipped with a light indicator which changes color and which allows the machine operating status to be recognized even from a distance. The indicator light identifies the statuses: Press ready, Molding cycle in progress, Temperature in the adjustment phase.

Moulds 
Moulds for producing plates and different types of samples according to international standards are available. Fixing rails with sliding system permit easily handling of the moulds. Mould produced upon customer design and company logo engraving are available on request.



Mould Extraction 
• All the mold models made by Gibitre can be applied to the press using clamping guides.
• The sliding guide on the lower platen allows you to move the mold effortlessly to facilitate the operations of inserting the product to be moulded and extracting the finished moulded part.


Standard Calibration service for Laboratory Press 
Maintenance and Calibration service for your Laboratory Press.
The service includes:
• Ordinaly maintenance of the instrument
• Calibration of the temperature of the platens. The temperature is measured, with the plates closed, in 9 different points of the platens and at 3 different temperatures (140°C, 170°C, 200°C).
• Calibration of the closure Force (50, 100, 150, 250 kN)
• Issue and e-mail shipment of the Calibration Certificate with traceability to primary standards.


Safety devices 
• Safety protection doors fitted with safety switch.
• Safety Push-button
• CE labelling




Dimension & Layout 


Development and production 
The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.



Instrument Characteristics 
 
Max useable platen surface 
270 mm x 270 mm 
Max closure force 
25 Tons 
Max platen stroke 
200 mm 
Temperatures 
Up to 250°C (300°C optional).
0.1°C resolution.
Maximimum error at the centre of the platen <0,5°C.
Mean difference in the 200x200mm central area of the platen from the center <0,5°C. 
Cooling System by Water Circuit (Option) 
Cooling of high efficiency test plates with controlled cooling gradient (between 0 and -50 °C/min) 
Cooling System by Compressed Air circuit (Option) 
The device is designed to accelerate the preparation of the press between moldings at different temperatures (Cooling speed 3°C/min) 
Closure force Control (option) 
Press configuration for continuous control of the closing force between 10 and 250 kN (between 1000 and 25000 kg) with maximum adjustment error 1 kN. Closure force can be set for each step of the molding cycle. 
Press control device 
 
PLC with Touch Screen Display 
The 10" color display permits to manually operate the machine, view the status of all the devices and sensors installed and prepare moulding cycles. 
Setup of Moulding cycles 
Storage of 40 molding cycles. Each cycle can include up to 30 operations which include: platen displacement, temperature set closure force set (option). 
Operating Status Indicator Light: 
Press ready, Moulding Cycle active, Press setting test temperature 
Safety Devices 
 
Safety Devices 
• Safety Pushbutton.
• Pneumatic safety protection Door with safety switch 


Labelling 
CE Labelling 
Calibration 
 
Calibration Report 
Calibration report with traceability to primary standards in conformity with the Calibration requirements specified in ISO 2393 standard 
Calibrated parameters 
The certificate includes the calibration of:
- temperature of the platens. The measure is performed in 9 different points of the platens and at 3 different temperatures (140°C, 170°C, 200°C).
- Closure force (with Pressure control option) 
Construction Characteristics 
 
Platen temperature control 
- Electrical heating.
- Independent temperature control of the plates by means of PID controllers 
Compressed Air supply 
Pressure: 6 bar. Compressed air regulation unit integrated into the instrument 
Power supply 
380 VAC ±10%,50 Hz ±3, 8 kW,three-phase -Other on request 
Electrical power 
8 kW 
Instrument dimensions 
(Width x Depth x Height) 872 x 1123 x 1943 mm 
Weight 
700 Kg 
Cooling System 
 
Cooling System 
allows the temperature reduction of the platens at a controlled speed (up to -50 ° C / min). The cooling circuit can be connected to a demineralised water supply source or to a chiller. 
Cooling liquid supply 
The water supplied to the Press must be demineralized.
Water pressure between 4 and 6 bar
Nominal flow rate: 0.3 L/sec at 25°C (standard water supply)  


Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Programmable Laboratory press for compression moulding, with 270x270 mm platens and 25 T closure force
Campi di applicazione  >  Prove su pneumatici  >  Rheocheck MD - Drive
Laboratory
Press -
Drive
lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。机器全貌。
Get a Quotation


Laboratory Press - Drive - 用于压缩成型的可编程实验室压机,具有 270x270 mm 的压板和 25 T 的闭合力
符合国际标准的实验室压片机,用于制备板材和样品。该设备经过专门设计,确保加热表面的热均匀性和机械强度,以保证制备出的试样具有一致的特性和厚度。
主要特点
• 机械结构采用4根立柱(直径60毫米)及由自润滑轴承导向的滑动板
• PLC控制系统配备10英寸触摸屏面板,实现对设备的全面控制及40个自动成型周期的设置
• 平面加热元件,专为确保板表面温度均匀分布而设计
• 气动控制的成型区域防护门,配备安全联锁装置和玻璃表面
• 成型区域封闭,便于烟雾排放
• CE认证
配件
• 冷却系统:控制板的冷却速度,适用于热塑性产品的成型
• 闭合力控制:允许在成型循环的每个步骤中连续调整闭合力
• Press_Control 软件:允许存储成型循环、准备待成型产品清单、根据产品自动调整压机、将成型曲线存储到 SQL 数据库
• 模具:用于生产板材及符合国际标准或特定客户要求的各种样品。

Programmable Laboratory press for compression moulding, with 270x270 mm platens and 25 T closure force

Applications

Laboratory press for preparing plates and test specimens compliant with international standards, engineered for uniform heat distribution on the platens and high mechanical rigidity. The result is specimens with consistent properties and thickness, ready for measurement tools and materials quality control.

Ideal for producing specimens to established practices, including compression of thermoplastics (ISO 293, ASTM D1928, ASTM D4703-16) and elastomeric compounds (ISO 2393, ASTM D3182), as well as plates for tests such as ISO 16770 and ISO 6916-1.

A four-column frame with self-lubricating bushings ensures precise alignment and stability; flat heating elements maintain even temperature. A PLC with 10'' touchscreen manages up to 40 automatic cycles, boosting repeatability and productivity. Safety and cleanliness are ensured by an interlocked pneumatic door and enclosed area with fume extraction, CE marked.

Accessories broaden applications: controlled cooling for thermoplastics, phase-b
Press to Start & Start to Press

Overview 

The laboratory press is an important tool for the preparing of samples with reproducible characteristics.
Uniform temperature distribution over the platen Surface and mechanical solidity to ensure constant thickness of the sample and consistent closure force are basic elements to ensure this result.

lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。带有开放式保护板的工作区细节,用于装卸试样。
用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。带有开放式保护板的工作区细节,用于装卸试样。

Key Features 

• Developed and Produced in Italy
• Fully conforming to ISO 2393, ISO 293, ASTM D3182 Standards
• Mechanical structure with 4 columns (with 60 mm diameter) 
• Hydraulic system with 25 tons closure force
• 270x270 mm chrome plated steel platens 
• Extremely uniform temperature distribution over the platen surface (± 0.5°C)
• Hi speed heating (up to 15°C/min) and Cooling (up to 50°C/min)
• Touch-Screen display for complete machine control
• 40 programmable moulding cycles
• Connection to software for storage to SQL database of the moulding cycles & curves of your products
• Closed moulding area for easy fume aspiration
• Pneumatic Protection doors with safety lock
• CE Labelling

lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。带有封闭保护板的工作区细节。 lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。带有封闭保护板的工作区细节。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。带活动保护板的工作区视图。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 驱动型实验室压力机:联锁气动开门,保护工作区。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。带活动保护板的工作区视图。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 实验室压力机 - 驱动装置:保护工作区的气动门细节。 lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。带有封闭保护板的工作区细节。

Tiltable Press Control Display 

The control of all the machine functions is carried out through the Touch-Screen Display integrated in the front part.
The 10" color display has been developed to ensure ergonomic use by the operator: it can be tilted during programming operations and positioned in a vertical position so as not to hinder access to the molding area.

lab press: 驱动型实验室压力机。用于机器控制的倾斜显示器。 lab press: 驱动型实验室压力机。用于机器控制的倾斜显示器。 lab press: 实验室-新闻:触摸显示屏 lab press: 实验室-新闻:触摸显示屏 lab press: 实验室-新闻:触摸显示屏 lab press: 驱动型实验室压力机。用于机器控制的倾斜显示器。 lab press: 用于模制橡胶和塑料试样的实验室压力机。机器的布局和尺寸。侧视图。 lab press: 用于模制橡胶和塑料试样的实验室压力机。机器的布局和尺寸。侧视图。 lab press: 用于模制橡胶和塑料试样的实验室压力机。机器的布局和尺寸。侧视图。
lab press: 实验室压榨机:压榨机控制面板细节。
 
Total Control of the Moulding Cycle

Press Control Functions 

The PLC controller for the press is provided with a user-friendly Touch Screen Display and permits the free definition of the moulding cycles by setting:
• The displacements of the moving platen of the press,
• The thermal cycle (different temperatures of the platens and thermal gradients can be set within a moulding cycle)
• The pressure cycle (with the force control option installed).
40 Moulding cycles can be stored. Each cycle includes 30 moulding steps.

lab press: 实验室压力机。控制 PLC。触摸屏控制面板主页。 lab press: 实验室压力机。控制 PLC。触摸屏控制面板主页。 lab press: 实验室压力机。控制 PLC。可储存和选择成型周期列表。 lab press: 实验室压力机。控制 PLC。用于自动执行的存储测试周期详情。 lab press: 实验室压力机。控制 PLC。带速度控制冷却示例的主显示页面。 lab press: Laboratory-Press-Drive (实验室-按压-驱动):主页面允许您激活选定的测试周期。 lab press: 实验室压力机。控制 PLC。带速度控制加热示例的主显示页面。

Gibitre_Press_Control Software: complete traceability of the sample preparation 

The press control PLC can be connected (via USB connection) to the Gibitre_Press_Control software.
The software allows you to:
Prepare the list of products to be moulded
• Automatically send the Pre-Set temperature, for the product to be moulded, to the PLC of the press and activate its automatic adjustment
• Automatically send the test cycle required, for the product to be moulded, to the press PLC
• Display the trend curves of the platen temperatures and the closing force (if the option is installed) during the execution of the molding cycle.
Store curves and numerical data in the Gibitre SQL database to keep the complete traceability of the specimen preparation cycle.
Gibitre_Press_Control software is compatible with Windows 10 and 11.

lab press: 用于控制实验室压力机的软件。主页。 lab press: 用于控制实验室压力机的软件。主页。 lab press: 实验室压机数据记录软件可存储每次成型循环的温度曲线和闭合力曲线。Datagest软件页面支持选择并显示测试结果。

Closure Force Control 

The Digital pressure regulation system is designed for applications that require an accurate control of closure force.
A specific 2-stages hydraulic circuit, which includes proportional regulation valves and continuous fine regulation pumping system, has been designed to ensure accurate pressure control.
The results is a smooth and consistent force control which permit to set the closure force between 10 kN and 250 kN with 10 N set force regulation steps with regulation oscillations lower than 1000 N.
The control of the closing force is performed using the Automatic Press Control that must be installed in combination with the Closure Force Control option.
The Automatic Press Control device permits to set multiple pressure regulation steps within the moulding cycle.

lab press: 实验室压力机。控制 PLC。不同温度和锁模力下的成型周期设置结果示例。 lab press: 实验室压力机。控制 PLC。不同温度和锁模力下的成型周期设置结果示例。 lab press: 实验室压榨机:具有多个压力和温度控制步骤的循环示例。 lab press: 实验室压力机。控制 PLC。用于自动执行的存储测试周期详情。
lab press: 实验室压机驱动版。控制工作台温度。使用热成像仪测量温度。
Temperature under control

Temperarture of the Platens 

The heating of the press surfaces is carried out by means of flat heating elements specially designed to meet the requirements of the ISO 2393 standard which admits a maximum difference of ± 0.5 °C inside the surface of the platens.
The temperature calibration is carried out through a rigorous control procedure which involves the measurement of 10 points distributed within the molding area.

The heating elements, with a power of 3.5 kW each, allow you to set heating gradients of the platens up to 15 °C/min.
Note: the maximum speed is conditioned by the thermal mass of the mold and the material to be molded.

Look at the Movie
lab press: 实验室印刷机。用热像仪拍摄的图像,以突出表面的热均匀性。
实验室印刷机。用热像仪拍摄的图像,以突出表面的热均匀性。

Cooling system with Water circuit 

The cooling of the press platens during the molding cycle is necessary for the molding of thermoplastic materials because the material must be cooled down to the solidification temperature before opening the mold.

Gibitre presses can be equipped with an efficient cooling device that allows the temperature reduction of the platens at a controlled speed (down to -50°C/min).

The cooling device consists of a cooling circuit that can be connected to a demineralised water supply source or to a chiller.

The cooling speed is set in the test cycle and is modulated by the control PID independently in the two planes of the press.

lab press: 用于储存冷却水的隔热箱 lab press: 用于储存冷却水的隔热箱 lab press: 实验室压力机。控制 PLC。带速度控制冷却示例的主显示页面。

Cooling system with Compressed Air circuit 

Gibitre presses can be equipped with a cooling device by means of compressed air which allows the temperature reduction of the platens at a controlled speed (down to -3 °C/min).

The device allows to speed up the regulation of the press temperature when molding operations at different temperatures have to be carried out.

The cooling speed is set in the test cycle and is modulated by the control PID independently in the two planes of the press.

lab press: 实验室压力机。控制 PLC。带速度控制冷却示例的主显示页面。
实验室压力机。控制 PLC。带速度控制冷却示例的主显示页面。
lab press: Gibitre 实验室印刷机:滑模处理细节。
Designed for the user

Light panel 

The press is equipped with a light indicator which changes color and which allows the machine operating status to be recognized even from a distance. The indicator light identifies the statuses: Press ready, Molding cycle in progress, Temperature in the adjustment phase.

lab press: 远距离检查仪器状态 lab press: 远距离检查仪器状态 lab press: 驱动型仪表:仪表上的前部发光标识显示机器的运行状态。 lab press: 驱动型仪表:仪表上的前部发光标识显示机器的运行状态。 lab press: 驱动型仪表:仪表上的前部发光标识显示机器的运行状态。 lab press: 驱动型仪表:仪表上的前部发光标识显示机器的运行状态。 lab press: 驱动型仪表:仪表上的前部发光标识显示机器的运行状态。 lab press: 驱动型仪表:仪表上的前部发光标识显示机器的运行状态。

Moulds 

Moulds for producing plates and different types of samples according to international standards are available. Fixing rails with sliding system permit easily handling of the moulds. Mould produced upon customer design and company logo engraving are available on request.

Read more about Gibitre Moulds
lab press: 用于制备 16 邵氏硬度试样的实验室压模。
用于制备 16 邵氏硬度试样的实验室压模。

Mould Extraction 

• All the mold models made by Gibitre can be applied to the press using clamping guides.
The sliding guide on the lower platen allows you to move the mold effortlessly to facilitate the operations of inserting the product to be moulded and extracting the finished moulded part.

lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。工作区细节和通过滑动导轨提取模具。 lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。工作区细节和通过滑动导轨提取模具。 lab press: 实验室压力机 - 驱动:通过滑动导轨移动下模。 lab press: 驱动型实验室压力机:通过滑动导轨实现下模移动。 lab press: 实验室压力机 - 驱动:通过滑动导轨移动下模。 lab press: 实验室压力机 - 驱动:通过滑动导轨移动下模。 lab press: 实验室压力机 - 驱动:通过滑动导轨移动下模。 lab press: 驱动型实验室压力机:通过滑动导轨实现下模移动。 lab press: 实验室压力机 - 驱动:通过滑动导轨移动下模。 lab press: 用于按照 ISO 293 和 ISO 2393 等标准制备试样的实验室压力机。工作区细节和通过滑动导轨提取模具。
Safety first

Standard Calibration service for Laboratory Press 

Maintenance and Calibration service for your Laboratory Press.
The service includes:
• Ordinaly maintenance of the instrument
• Calibration of the temperature of the platens. The temperature is measured, with the plates closed, in 9 different points of the platens and at 3 different temperatures (140°C, 170°C, 200°C).
• Calibration of the closure Force (50, 100, 150, 250 kN)
• Issue and e-mail shipment of the Calibration Certificate with traceability to primary standards.

Sample Calibration Report
lab press: 吉比特校准:完全符合标准
吉比特校准:完全符合标准

Safety devices 

• Safety protection doors fitted with safety switch.
• Safety Push-button
• CE labelling

lab press: 安全第一
安全第一

Dimension & Layout 

lab press: 用于模制橡胶和塑料试样的实验室压力机。机器的布局和尺寸。侧视图。 lab press: 用于模制橡胶和塑料试样的实验室压力机。机器的布局和尺寸。侧视图。 lab press: 实验室压力机:机器尺寸。侧视图和正视图。 lab press: 实验室压力机:机器尺寸,典型安装配置,可选配压盘冷却装置。
Proudly Made in Italy

Development and production 

The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.

lab press: 引以为豪的意大利制造
引以为豪的意大利制造

Technical specification

Reference Standards
Standards the instrument complies with:  EN 205; ISO 6916-2; ISO 293; ISO 16770; ASTM D1928; ASTM D4703-16; ISO 2393; ASTM D3182; ISO 6916-1; UNI 5572;  

Instrument Characteristics

Max useable platen surface  270 mm x 270 mm  Max closure force  25 Tons  Max platen stroke  200 mm  Temperatures  Up to 250°C (300°C optional).
0.1°C resolution.
Maximimum error at the centre of the platen <0,5°C.
Mean difference in the 200x200mm central area of the platen from the center <0,5°C. 
Cooling System by Water Circuit (Option)  Cooling of high efficiency test plates with controlled cooling gradient (between 0 and -50 °C/min)  Cooling System by Compressed Air circuit (Option)  The device is designed to accelerate the preparation of the press between moldings at different temperatures (Cooling speed 3°C/min)  Closure force Control (option)  Press configuration for continuous control of the closing force between 10 and 250 kN (between 1000 and 25000 kg) with maximum adjustment error 1 kN. Closure force can be set for each step of the molding cycle. 

Press control device

PLC with Touch Screen Display  The 10" color display permits to manually operate the machine, view the status of all the devices and sensors installed and prepare moulding cycles.  Setup of Moulding cycles  Storage of 40 molding cycles. Each cycle can include up to 30 operations which include: platen displacement, temperature set closure force set (option).  Operating Status Indicator Light:  Press ready, Moulding Cycle active, Press setting test temperature 

Safety Devices

Safety Devices  • Safety Pushbutton.
• Pneumatic safety protection Door with safety switch 
Labelling  CE Labelling 

Calibration

Calibration Report  Calibration report with traceability to primary standards in conformity with the Calibration requirements specified in ISO 2393 standard  Calibrated parameters  The certificate includes the calibration of:
- temperature of the platens. The measure is performed in 9 different points of the platens and at 3 different temperatures (140°C, 170°C, 200°C).
- Closure force (with Pressure control option) 

Construction Characteristics

Platen temperature control  - Electrical heating.
- Independent temperature control of the plates by means of PID controllers 
Compressed Air supply  Pressure: 6 bar. Compressed air regulation unit integrated into the instrument  Power supply  380 VAC ±10%,50 Hz ±3, 8 kW,three-phase -Other on request  Electrical power  8 kW  Instrument dimensions  (Width x Depth x Height) 872 x 1123 x 1943 mm  Weight  700 Kg 

Cooling System

Cooling System  allows the temperature reduction of the platens at a controlled speed (up to -50 ° C / min). The cooling circuit can be connected to a demineralised water supply source or to a chiller.  Cooling liquid supply  The water supplied to the Press must be demineralized.
Water pressure between 4 and 6 bar
Nominal flow rate: 0.3 L/sec at 25°C (standard water supply)  

FAQ

References
Gibitre Laboratory Press
is currently installed in
42 countries
Request the List of top References
Download
Brief product leaflet
General catalogue mail request

Calibration certificate
example
CE certificate
example
Moulds for Laboratory Press
Vedi anche: 实验室冲压模具:磨光镀铬模具示例。
See Also   >
Multi-Head Die Cutter
Vedi anche: 手动模切机,用于使用模具制备试样。
See Also   >
News
Call to action
Ask for quotation
a guided way to ask
for a quotation
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
We welcome your calls Monday through Friday from 8:30 a.m. to 5:30 p.m. in your local time zone.
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Gibitre Instruments S.r.l., Via dell'industria 73, 24126 Bergamo Italia, pec@gibitre-instruments.it P. IVA: IT02827900164 CCIAA REA Bergamo: 326486 | ©2018 - 2026 Gibitre Instruments
Optimist.it
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.