Standards the instrument complies with:   
ASTM D1149;   ISO 1431-1;    

Support for dynamic test 
Device, in compliance with ISO 1431, ASTM D1149 and SAE J1401, to be inserted into test chamber for the study of ozone resistance of cured rubber under dynamic conditions (from 3 to 30 cycles/min - 0.05 to 0.5 Hz).
Up to 10 samples can be tested at the same time


Control Display 
The instrument's Touch-Screen Display allows you to set:
• The number of swings to be made
• The Oscillation frequency

The instrument will stop automatically when the set number of oscillations is reached


Adjustment of instrument displacement 
The instrument allows the mechanical adjustment of:
• the distance between the clamps for fixing or specimens (default 50 mm)
• the stroke performed by the mobile clamp: from 2.5 to 30 mm (between 5% and 60% for 50 mm specimen in accordance with the standards)




Insertion into the Test Chamber 
The Dynamic Tester is inserted into the test chamber using the motorized winch.
The same winch allows you to lift the device to check the condition of the specimens.

The lifting winch is controlled by the interlocked safety device which inhibits its operation if the ozone concentration inside the chamber is above the safety limits (10 pphm)



Number of Samples 
10 
Frequency 
3 to 30 cicles/min (0.05 to 0.5 Hz) 
Adjustment of Displacement 
2.5 to 30 mm (from 5% to 60% for standard 50 mm Sample) 


Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Gibitre Instruments

Support for dynamic test in Ozone

Support for the execution of Dynamic Tests within the OzoneCheck Instrument




 
Support for dynamic test 
Device, in compliance with ISO 1431, ASTM D1149 and SAE J1401, to be inserted into test chamber for the study of ozone resistance of cured rubber under dynamic conditions (from 3 to 30 cycles/min - 0.05 to 0.5 Hz).
Up to 10 samples can be tested at the same time

Control Display 
The instrument's Touch-Screen Display allows you to set:
• The number of swings to be made
• The Oscillation frequency

The instrument will stop automatically when the set number of oscillations is reached

Adjustment of instrument displacement 
The instrument allows the mechanical adjustment of:
• the distance between the clamps for fixing or specimens (default 50 mm)
• the stroke performed by the mobile clamp: from 2.5 to 30 mm (between 5% and 60% for 50 mm specimen in accordance with the standards)
 
Insertion into the Test Chamber 
The Dynamic Tester is inserted into the test chamber using the motorized winch.
The same winch allows you to lift the device to check the condition of the specimens.

The lifting winch is controlled by the interlocked safety device which inhibits its operation if the ozone concentration inside the chamber is above the safety limits (10 pphm)

 
Technical specifications
Reference Standards
Standards the instrument complies with:  ASTM D1149; ISO 1431-1;  
Number of Samples  10  Frequency  3 to 30 cicles/min (0.05 to 0.5 Hz)  Adjustment of Displacement  2.5 to 30 mm (from 5% to 60% for standard 50 mm Sample) 
Download
General catalogue

 
 
Ask for quotation

Ask for quotation
Mail

Mail
Mail
Phone

Phone
Phone
Local agent


Local agent
FAQ
do you need more information or a quotation?
use the buttons at the bottom of the page.
Cookie policy
Questo sito utilizza i cookies (dati permanenti registrati sul tuo PC) per:
- ricordare la lingua scelta
- ricordare i prodotti preferiti
- suggerire prodotti in funzione delle pagine più visitate
- mantenere aperta la sessione (nella sole aree riservata)

Cliccando su "accetta" autorizzate l'uso dei cookies.

Accetto / Accept

This site uses cookies (permanent data on your PC) for:
- remembering the choosed language
- remembering your favorite products
- suggest other products based upon the most viewed pages
- keep the navigation session open (only in the reserved area)

By ckicking on "accept" you authorize the use of cookies.