Compression Set - Adjustable
Compression Set device made of Stainless Steel with adjustable distance between the plates


Standards the instrument complies with:   
ASTM D1414;   ASTM D395-B;   FIAT 50410;   ISO 815;   ISO 1856;    

 
The Gibitre Instruments testing tool is engineered for finished articles. The plate gap adapts to specimen thickness and required strain, accommodating different materials and delivering consistent measurement on O‑rings, small parts, and standard specimens. All‑stainless steel construction, a 100 mm usable area, and precise surface parallelism (flatness error <0.02 mm) ensure uniform compression and reliable results, supporting testing to ISO 815, ASTM D1414, and ASTM D395‑B.

Why choose it: the distance between test surfaces is adjustable from 0 to 20 mm, offering flexibility without sacrificing accuracy. The integrated thickness gauge, with 0.001 mm resolution and compliant with relevant standards, streamlines plate setup and measurement result calculation. Dedicated software acquires feeler‑gauge data and calculates compression/set, making testing fast, repeatable, and traceable. A complete solution for those who need essential, accurate tools for sample and materials characterizat

Overview 
The instrument is designed for the execution of tests on finished articles because the distance of the plates can be adjusted according to the thickness of the samples and of the deformation to be applied.
The accurate construction method makes it ideal both for the testing on O-rings or small-size finished articles and for standard specimens.
Key Features
• Entirely made of stainless steel
• Wide usable surface (100 mm diameter)
• Adjustable distance between the test surfaces between 0 and 20 mm
• Perfect parallelism between the test surfaces (flatness error less than 0.01mm)

The adjustment of the distance between the plates is performed using a standard thickness meter (optional).


Use of the instrument 
Performing a test is easy:
• lock the lower part of the device in a vice
• insert and lock the dial gauge in the hole of the upper plane
• turn the adjusting ring to set the distance between test planes to zero
• set the dial gauge to zero
• measure the thickness of the specimen and calculate the distance between the planes necessary to obtain the required deformation during the test
• adjust the distance between the planes by means of the adjustment ring (the dial gauge shows the real distance between the plates)




 
• lock the position of the surfaces using the locking ring nut
• remove the dial gauge
• remove the upper table to insert the specimens and close it again

You device is ready! Inserted it in the oven for the required time.
After aging in the oven, you only need to remove the top and measure the final thickness of the specimens.

NOTE: further tests on pieces with the same thickness don't require any additional adjustment.


Surface Rugosity of the plates 
• The measurement of the surface roughness of the device is reported in the Calibration Report of the instrument and guarantees full compliance with the requirements of the standards.
• The Identification Code imprinted on the Plates guarantees the traceability of the measurements made.


Accessories 
Moulds for preparing samples in compliance with standards.


Testing Configuration 
 
Thickness regulation 
Between 0 and 20 mm.
Adjustment is carried out using a comparator (optional). 
Usable test area 
Diameter 100 mm 
Technical specifications 
 
External Dimensions 
Diameter 140 mm, Height 120 mm 
Construction Characteristics 
18 mm thick chrome-plated steel shelves suitable for use up to 250°C 
Weight 
8 kg 
Calibration 
Calibration report with traceability to the instruments used. The Calibration includes the Measurement of surface roughness of the test plates 
Roughness of the plates 
Ra < 0.4 µm  


Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Compression Set device made of Stainless Steel with adjustable distance between the plates
Campi di applicazione  >  Prove su pneumatici  >  Rheocheck MD - Drive
Compression
Set - Adjustable
compression set adjustable: Dispositivo per compression set per articoli tecnici con distanza regolabile tra i piani.
Get a Quotation

Compression Set device made of Stainless Steel with adjustable distance between the plates

Applications

The Gibitre Instruments testing tool is engineered for finished articles. The plate gap adapts to specimen thickness and required strain, accommodating different materials and delivering consistent measurement on O‑rings, small parts, and standard specimens. All‑stainless steel construction, a 100 mm usable area, and precise surface parallelism (flatness error <0.02 mm) ensure uniform compression and reliable results, supporting testing to ISO 815, ASTM D1414, and ASTM D395‑B.

Why choose it: the distance between test surfaces is adjustable from 0 to 20 mm, offering flexibility without sacrificing accuracy. The integrated thickness gauge, with 0.001 mm resolution and compliant with relevant standards, streamlines plate setup and measurement result calculation. Dedicated software acquires feeler‑gauge data and calculates compression/set, making testing fast, repeatable, and traceable. A complete solution for those who need essential, accurate tools for sample and materials characterizat
Perché scegliere Compression Set Regolabile ?

Overview 

The instrument is designed for the execution of tests on finished articles because the distance of the plates can be adjusted according to the thickness of the samples and of the deformation to be applied.
The accurate construction method makes it ideal both for the testing on O-rings or small-size finished articles and for standard specimens.
Key Features
• Entirely made of stainless steel
• Wide usable surface (100 mm diameter)
• Adjustable distance between the test surfaces between 0 and 20 mm
• Perfect parallelism between the test surfaces (flatness error less than 0.01mm)

The adjustment of the distance between the plates is performed using a standard thickness meter (optional).

compression set adjustable: Dispositivo di compression set con distanza regolabile dei piani. Esempio di prova su o-ring di piccolo diametro. compression set adjustable: Dispositivo di compression set con distanza regolabile dei piani. Esempio di prova su o-ring di piccolo diametro. compression set adjustable: Dispositivo per compression set con distanza regolabile tra i piani. Esempio di test su o-ring. compression set adjustable: Dispositivo per compression set con distanza regolabile tra i piani. Esempio di test su o-ring.

Use of the instrument 

Performing a test is easy:
• lock the lower part of the device in a vice
• insert and lock the dial gauge in the hole of the upper plane
• turn the adjusting ring to set the distance between test planes to zero
• set the dial gauge to zero
• measure the thickness of the specimen and calculate the distance between the planes necessary to obtain the required deformation during the test
• adjust the distance between the planes by means of the adjustment ring (the dial gauge shows the real distance between the plates)

compression set adjustable: Prove di compression set. Dispositivo con distanza piani regolabile. Regolazione della distanza tra i piani con comparatore millesimale rimovibile. compression set adjustable: Prove di compression set. Dispositivo con distanza piani regolabile. Regolazione della distanza tra i piani con comparatore millesimale rimovibile. compression set adjustable: Dispositivo per compression set con distanza regolabile tra i piani. Esempio di test su o-ring. compression set adjustable: Dispositivo per compression set con distanza regolabile tra i piani. Regolazione della distanza.

 

• lock the position of the surfaces using the locking ring nut
• remove the dial gauge
• remove the upper table to insert the specimens and close it again

You device is ready! Inserted it in the oven for the required time.
After aging in the oven, you only need to remove the top and measure the final thickness of the specimens.

NOTE: further tests on pieces with the same thickness don't require any additional adjustment.

compression set adjustable: Compression set regolabile. Chiusura del dispositivo dopo l'inserimento dell'articolo da provare. compression set adjustable: Compression set regolabile. Chiusura del dispositivo dopo l'inserimento dell'articolo da provare. compression set adjustable: Dispositivo per compression set con distanza regolabile tra i piani. Preparazione del dispositivo.
 

Surface Rugosity of the plates 

• The measurement of the surface roughness of the device is reported in the Calibration Report of the instrument and guarantees full compliance with the requirements of the standards.
• The Identification Code imprinted on the Plates guarantees the traceability of the measurements made.

compression set adjustable: Dispositivo per compression set con distanza regolabile. Certificato di taratura della rugosità superficiale dei piani. compression set adjustable: Dispositivo per compression set con distanza regolabile. Certificato di taratura della rugosità superficiale dei piani. compression set adjustable: Dispositivo per compression set. Taratura della rugosità delle piastre in conformità alla norma ISO 815. compression set adjustable: Dispositivo per compression set. Taratura della rugosità delle piastre in accordo a norma ASTM D395.

Accessories 

Moulds for preparing samples in compliance with standards.

compression set adjustable: Stampo in tre parti per la preparazione di provini per prove di compression set. compression set adjustable: Stampo in tre parti per la preparazione di provini per prove di compression set. compression set adjustable: Stampi per pressa da laboratorio. Stampo per provini di compression set con base sorrevole e parte centrale mobile per l'estrazione di provini. compression set adjustable: Stampi per pressa. Stampo con base sorrevole ed inserto centrale intercambiabile. Nella foto l'inserto per la preparazione di provini per abrasione.

Technical specification

Reference Standards
Standards the instrument complies with:  FIAT 50410; ISO 1856; ASTM D1414; ISO 815; ASTM D395-B;  

Testing Configuration

Thickness regulation  Between 0 and 20 mm.
Adjustment is carried out using a comparator (optional). 
Usable test area  Diameter 100 mm 

Technical specifications

External Dimensions  Diameter 140 mm, Height 120 mm  Construction Characteristics  18 mm thick chrome-plated steel shelves suitable for use up to 250°C  Weight  8 kg  Calibration  Calibration report with traceability to the instruments used. The Calibration includes the Measurement of surface roughness of the test plates  Roughness of the plates  Ra < 0.4 µm  
Download
Brief product leaflet
General catalogue mail request

Thickness meter with adjustable support
Vedi anche: Misurazione spessore provini: il comparatore Mitutoyo con supporto conforme alle norme.
See Also   >
Compression_Set - Software
Vedi anche: Software Gibitre. Software per prova di compression set integrato con comparatore. Pagina di inserimento delle misure.
See Also   >
Compression Set - Drive
Vedi anche: Nuovo modello di dispositivo per prove di Compression set conforme alle norme ISO 815, ASTM D395 Metodo B, DIN 53 517.
See Also   >
News
Call to action
Ask for quotation
a guided way to ask
for a quotation
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
We welcome your calls Monday through Friday from 8:30 a.m. to 5:30 p.m. in your local time zone.
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Gibitre Instruments S.r.l., Via dell'industria 73, 24126 Bergamo Italia, pec@gibitre-instruments.it P. IVA: IT02827900164 CCIAA REA Bergamo: 326486 | ©2018 - 2026 Gibitre Instruments
Optimist.it
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.