Datagest
Software
Gibitre Software
Platform
Smart Test
Identification
RheoCheck
Software
Mooney
Check - Software
Density
Check - Software
Hardness
Check - Software
Tensor
Check - Software
TR_Test -
Software
Brittleness
Check -
Software
PressCheck
Software
Compression
Set -
Software
Company Connection
Software
Gibitre
Software Installer
Maintenance and Calibration Service
ISO 17025 Accredited Calibration
Remote Assistance Service
Support request form



















































































































































































































































































































































Italiano
English
Español
Français
Deutsch
Português
Polskie
Čeština
Magyar
Русский
Türk
中文
日本人
한국어
Arab











































The calculation returns the Mooney Viscosity after the set test time. Standard test time includes 1 minute pre-heating time and 4 minutes test time.
REFERENCE STANDARD: ISO 289-1 / ASTM D 1646
UNIT: MU (Mooney units)
CALCULATION PARAMETERS:
Test Time: t
EXPRESSION:
ML (1+t) - Example ML (1+4)

The calculation returns the minimum value of Mooney Viscosity over the complete test time.
UNIT: MU (Mooney units)
EXPRESSION:
MU_min

The calculation returns the Maximum Mooney viscosity at the beginning of test time.
UNIT: MU (Mooney units)
EXPRESSION:
MU_ini


The calculation returns the Slope of curve of the Viscosity vs Time at the set test time.
UNIT: MU (Mooney units)
PARAMETERS:
Test Time: t
EXPRESSION:
S (1+t) - Example S(1+12)

Reference standard ISO 289-3
The calculation returns the difference between the Mooney viscosities recorded at running time t1 and t2.
UNIT: MU (mooney units)
PARAMETERS:
Test Time t1
Test Time t2
EXPRESSION:
Dt_MU_(t2-t1)

OVERVIEW:
The pre-vulcanization characteristics provide a means of estimating how long compounded rubber can be maintained at high temperatures and remain processable.
REFERENCE STANDARD:
ISO 289-3 / ASTM D 1646
The calculation returns the time in minutes including warm-up time, for the viscosity to increase by a specified amount from the minimum value.
UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)
PARAMETERS:
Torque Increase: MU
EXPRESSION:
ts_MU (example ts_35)


The calculation returns the time needed for the viscosity to increase from the first to the second set values.
REFERENCE STANDARD: ASTM D 1646
UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)
PARAMETERS:
Start Torque Value: ts1
Final Torque Value: ts2
EXPRESSION:
Dt_ts_(ts2-ts1) - Example: Dt_ts(35-5)

OVERVIEW:
Stress Relaxation test consists of determining the decay of the Mooney torque immediately after the determination of the Mooney viscosity. After abruptly stopping the rotor at the end of the Mooney viscosity measurement, the decrease in torque is recorded as a function of time. The rate of change of the torque is evaluated over a short time interval.
REFERENCE STANDARD: ISO 289-4
CALCULATION:
The calculation returns the slope of the linear regression line of the log(torque) versus log(time) plot over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement. According to ISO Standard the regression is calculated between 1.6 and 5 seconds from rotor stop.
UNIT: Mu/sec
PARAMETERS:
Time for Start of Regression Calculation: t1
Time for Stop of Regression Calculation: t2
EXPRESSION:
a(MSR) t1_t2 - Example: a(MSR) 1.6_5.0


The calculation returns the area under the stress relaxation curve over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement.
REFERENCE STANDARD: ASTM D 1646
UNIT: Mu*sec
PARAMETERS:
Time for Start of the Calculation: t1
Time for Stop of the Calculation: t2
EXPRESSION:
A (t1_t2) - Example: A (1.6_5.0)