© Gibitre Instruments
P.IVA IT02827900164 Web Design Optimist
|
|
Ozone Check - Standard Version
Electronic Ozone Tester for the evaluation of ozone resistance of rubber samples
|
STANDARDS:
ISO 1431, 3011, 7326; ASTM D 1149, D 4575, D 3395, D 518; DIN 53 509
DOWNLOADS:
Leaflet »
Test Report »
|
TECHNICAL CHARACTERISTICS
|
Standards the instrument complies with
|
ISO 1431, 3011, 7326; ASTM D1149, D4575, D3395, D 518; DIN 53 509, SAE J1401
|
|
Test Data produced
|
Ozone concentration, temperature and Humidity level (only with Humidity control option) during test time
|
|
Data Printout
|
Test identification data, test parameters, sample characteristics, curves of test parameters and notes about results.
|
|
Temperature control
|
From Room Temperature + 5 °C to 70 °C by mean of microprocessor PID thermoregulators with 0.1 °C accuracy. The instrument is equipped with a heat interchanger that can be connected to a water supply to reduce the temperature.
|
|
Air flow
|
Adjustable from 0,5 to 3 changes/min (effective air change velocity between 7 and 39 mm/sec). Internal circulating fan according to ISO 1431.
|
|
Ozone concentration control
|
Automatic or manual between 10 and 500 PPHM by means of electrochemical Ozo-stop system with ± 3% accuracy
|
|
Ozone circuit type
|
Closed circuit with continuous ozone production and shooting down through self regenerating active carbon filter.
|
|
Static Sample Holders (optional)
|
- Sample holder for static strain of 2 samples with adjustable distance. - Test piece holder with planetary displacement in compliance with ASTM D 3395-A standard. - Sample holders for tests on rubber hoses according to ISO 7326-method 1 and 2.
|
|
Dynamic Holder (optional)
|
In compliance with ISO 1431 and SAE J1401 To be inserted into test chamber for the study of ozone resistance of cured rubber under dynamic conditions (from 3 to 30 cycles/min - 0.05 to 0.5 Hz). Up to 10 samples can be tested at the same time.
|
|
Test chamber
|
Cylindrical Stainless Steel test chamber with 155 litres capacity provided with adjustable internal illumination and large inspection window. Dimensions:diameter: 550 mm; height 550 mm
|
|
Sample holders Elevator
|
Lifting gear for up and down movement to help in inspection of the different sample holders
|
|
Power supply
|
220 VAC ±10%,50 Hz ±3, 10 A, single phase, 1.6 kW - Other on request
|
|
Dimensions
|
(W x D x H) 1060 x 1000 x 1080 (2390 with lifting system) mm
|
|
Weight
|
370 Kg
|
|
Personal computer
|
Min. Config.: Pentium IV 1 GB RAM.Oper.Systems: Windows NT, XP, Vista, 7 (64 bits) Connection to the instrument with USB port
|
|
Software usage Languages
|
Italian,English,French,Spanish,German,Portug.,Russian,Turkish,Chinese,Japan,Polish
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
.
|
|
|
|
|
|
|
|
Overview Gibitre Ozone Check is a complete ozone cabinet produced in conformity with international standards, which permits to estimate the resistance to
Technical features REGULATION OF THE INSTRUMENT The Ozone tester permits the regulation of: Ozone Concentration between 0 and 500 pphm; Test temperature between
Software
The program permits the preparation of customer-defined test procedures defining test parameters: ozone concentration, temperature, relative humidity
|
|
SOFTWARE:
Main Page Ozone Check Software The main page includes a complete description of the state of the devices of the instrument (Ozone generation device, air recirculation system, heating system, humidity generation system, safety lock). Complete information about the test status, the test Identification, and the curves of temperature, ozone concentration and humidity is immediately available.

|
|