TECHNICAL CHARACTERISTICS
|
Standards the instrument complies with
|
ISO 1431, 3011, 7326; ASTM D 1149, D 4575, D 3395, D 518; DIN 53 509
|
|
Test Data produced
|
Ozone concentration, temperature and Humidity level (only with Humidity control option) during test time
|
|
Data Printout
|
General test identification data Automatic or manual printout of ozone concentration and temperature during test time
|
|
Temperature control
|
From Room Temperature + 5 °C to 70 °C by mean of microprocessor PID thermoregulators with 0.1 °C accuracy. The instrument is equipped with a heat interchanger that can be connected to a water supply to reduce the temperature.
|
|
Air flow
|
Adjustable from 0,5 to 3 changes /min (effective air change velocity between 7 and 39 mm/sec). Internal circulating fan according to ISO 1431.
|
|
Ozone concentration control
|
Automatic or manual between 10 and 500 PPHM by means of electrochemical Ozo-stop system with ± 3% accuracy
|
|
Ozone circuit type
|
Closed circuit with continuous ozone production and shooting down through self regenerating active carbon filter.
|
|
Standard Sample Holder
|
Enables static tests to be carried out
|
|
Static Sample Holders (optional)
|
-Grips for static tests with sample traction -Static test piece holder with planetary effect which moves the test pieces over the whole volume of the chamber for a more homogeneous ozone atmosphere exposition.
|
|
Dynamic Holder (optional)
|
In compliance with ASTM D 3395-A standard. To be inserted into test chamber for the study of ozone resistance of cured rubber under dynamic conditions (from 3 to 30 cycles/min - 0.05 to 0.5 Hz). Up to 10 samples can be tested at the same time.
|
|
Test chamber
|
Cylindrical Stainless Steel test chamber with 155 litres capacity provided with adjustable internal illumination and large inspection window. Dimensions:diameter: 550 mm; height 550 mm
|
|
Sample holders Elevator
|
Lifting gear for up and down movement to help in inspection of the different sample holders
|
|
Power supply
|
220 VAC ±10%,50 Hz ±3, 10 A,single phase -Other on request
|
|
Power
|
1.6 Kw
|
|
Dimensions
|
(Width x Depth x Height) 1500 x 950 x 1200 mm
|
|
Weight
|
330 Kg
|
|
Personal computer
|
Minimum Configuration: Pentium IV 512 MB RAM Compatible operating Systems: Windows 2000, NT, XP, Vista Connection to the instrument with one serial or USB port
|
|
Software usage Languages
|
Italian, English, French, Spanish, German, Portuguese, Russian
|
|
|
|
|
|
|
|
|
.
|
|
Overview The Ozone Tester enables evaluation to be carried out of rubber ozone resistance of standard samples or technical articles in static and dynamic
Technical features REGULATION OF THE INSTRUMENT The Ozone tester enables regulation of the Ozone Concentration to be carried out between 0 and 500 pphm, the test
Software
The program permits the preparation of customer-defined test procedures defining test parameters: ozone concentration, temperature, relative humidity
|
The exclusive cylinder shaped test chamber permits a homogeneous ozone concentration inside the chamber
|
Planetary Effect device that permits to move 16 sample holders for static test in traction
|
Sample holder for 16 samples for Dynamic Tests according to ASTM D 3395-A
|
|
SOFTWARE:
Main Page Ozone Check Software The main page includes a complete description of the state of the devices of the instrument (Ozone generation device, air recirculation system, heating system, humidity generation system, safety lock). Complete information about the test status, the test Identification, and the curves of temperature, ozone concentration and humidity is immediately available.

|
|